Abstract:The influence of total ionic strength adjustment buffer addition on the determination of F- in solubility phase diagram of zinc hydrometallurgy system by fluorine ion selective electrode-standard curve method and standard addition method was systemically investigated. The research results showed that: (1) as a whole, the measurement results of fluorine ion selective electrode-standard addition method were much better than those obtained by standard curve method regardless of addition amount of total ionic strength adjustment buffer; (2) in fluorine ion selective electrode-standard addition method, the determination results were greatly affected by the addition amount of total ionic strength adjustment buffer; (3) when the background ion concentration in testing solution was very low, the addition of total ionic strength adjustment buffer had negative effect on the accuracy of analytical results. In other words, the results were better without addition of total ionic strength adjustment buffer; (4) when the background ion concentration in testing solution was relatively high, the addition of total ionic strength adjustment buffer was advantageous the improvement of determination precision. The experimental results were applied for the determination of F- in solution of zinc hydrometallurgy. The lowest error was 0.93%.
蔡德花, 曾德文. 离子缓冲剂对选择性电极法测定湿法炼锌体系中氟的影响[J]. 冶金分析, 2018, 38(5): 41-46.
CAI De-hua, ZENG De-wen. Influence of total ionic strength adjustment buffer on the determination of fluorine in zinc hydrometallurgy system by selective electrode method. , 2018, 38(5): 41-46.
William F,Linke A.Solubility,volume 1:Inorganic and Compound[M].American Chemical Society,1958.
[4]
Revinson D,Harley J H.Spectrophotometric determination of fluoride ion with chrome azurol S[J]. Analytical Chemistry,1953,25(5):794-797.
[5]
Hang Y P,Wu C Y.Ion chromatography for rapid and sensitive determination of fluoride in milk after headspace single-drop microextraction with in situ generation of volatile hydrogen fluoride[J].Analytica Chimica Acta,2010,661(2):161.
胡一航,王海北,王玉芳.水蒸气蒸馏-离子色谱法测定湿法炼锌溶液中氟和氯[J].冶金分析,2015,35(5):59-62.HU Yi-hang,WANG Hai-bei,WANG Yu-fang.Determination of fluorine and chlorine in solution of wet zinc hydrometallurgy by steam distillation-ion chromatography[J].Metallurgical Analysis,2015,35(5):59-62.
[8]
董惠英,胡正业.氟离子选择性电极法测定化工含氟废水[J].化工环保(Environmental Protection of Chemical Industry),1986(2):108-110,107.
Martell A E,Smith R M.Critical Stability constants,volume 4:inorganic complexes[M].NewYork:Plenum Press,1976.
[11]
韩威.用氟离子选择电极作湿法炼锌的控制分析[J].贵州大学学报(Journal of Guizhou University ),1975(2):37-42.
[12]
许康,张旭,沈庆峰,等.氟离子选择性电极-标准加入法测定硫酸锌溶液中氟[J].冶金分析,2014,34(5):41-45.XU Kang,ZHANG Xu,SHEN Qing-feng,et al.Standard addition method for the determination of fluoride in zinc sulfate solution by fluoride ion-selective electrode[J].Metallurgical Analysis,2014,34(5):41-45.