Abstract: Molybdenum ore and molybdenum concentrate were calcined by magnesium oxide. After hot water extraction, the sample was directly determined by inductively coupled plasma mass spectrometry to obtain the content of rhenium. A study was conducted on the calcination solvent & conditions, extraction conditions and instrumental determination conditions. Results indicated that: magnesium oxide was taken as calcination solvent; the sample was kept at 650 ℃ for 90 min; the optimal dissolution conditions was to use hot water for leaching; 103Rh was selected as internal element to effectively correct instrumental drift and reduce the interference of matrix effect. The detection limit was 0.005 μg/g and recoveries were between 94 % and 102 %. The method was tested by certified reference materials with the result consistent with certified value and RSD (n=5) ≤ 3.6 %. The method is suitable for the determination of rhenium with the content of 0.05~300 μg/g in molybdenum ore and molybdenum concentrate.
[6]王军芝, 杜安道. 硫化物矿物中铼和锇的ICP-MS测定—逆王水分解样品方法[J]. 质谱学报(Journal of Chinese Mass Spectrometry Society), 1994, 20(3,4): 111-112.
[7]Sun Y, Chu Z, Sun M, et al. An improved Fe-Ni sulfide fire assay method for determination of Re, Platinum group elements, and Os isotopic ratios by inductively coupled plasma- and negative thermal ionization-mass spectrometry [J].Applied Spectroscopy, 2009,63(11):1232-1237.
[8]Suzuki K, Lu Q, Shimizu H, et al. Determination of Osmium abundance in molybdenite mineral by isotope dilution mass spectrometry with microwave digestion using potassium dichromate as oxidizing agent[J]. Analyst, 1992,117(7):1151-1156.
[9]李永梅. 稀散元素铼的示波极谱分析法研究[J]. 河北能源职业技术学院学报(Journal of Hebei Energy Institute of Vocation and Technology), 2004, (1): 95-96.