Abstract:The sample was digested with 3 mL of sulfuric acid, 2 mL of hydrofluoric acid and 2 mL of hydrogen peroxide as solvent by 4-step temperature rising microwave digestion technique. A determination method of vanadium in titanium dioxide was established by graphite furnace atomic absorption spectrometry with ashing temperature of 1 500 °C and atomization temperature of 2 700 °C, respectively. The results showed that the absorbance of test solution could be effectively strengthened by adding 10 mL of 0.5 g/L magnesium nitrate solution as matrix modifier. The linear range of method was 0-50 μg/L, and the detection limit was 0.01 μg/g. This proposed method was applied to determine vanadium in titanium dioxide samples. The results were consistent with those obtained by inductively coupled plasma mass spectrometry after conventional wet digestion. The relative standard deviation (RSD, n=11) of determination results was 1.9%-3.2%, and the recoveries were between 98% and 100%.