Abstract:A microwave plasma atomic emission spectrometry(MP-AES) for the determination of Cu, Pb, Fe, Cd, Mn in zinc oxide produced by direct process with microwave plasma atomic emission spectrometer was established. The sample was dissolved with a mixtare of nitric acid and hydrochloric acid by heating at low temperature. The spectral lines of Cd 228.802 nm, Cu 324.754 nm, Fe 259.940 nm, Mn 403.076 nm and Pb 405.781 nm were selected as analytical lines. The matrix of zinc did not interfere in the determination of those elements within selected amount of sample. The relative standard deviation for the determination of each element in the sample of ZnO-1 was between 0.91% and 4.9%, and the standard addition recovery was between 90% and 116%. The determination results were consistent with those obtained by national standard and ICP-AES methods. The established method was suitable for simultaneous determination of multi-element in zinc oxide produced by direct process.
张士涛,章连香,冯先进. 电感耦合等离子体原子发射光谱法(ICP-OES)测定直接法氧化锌中铝、铜、铅、铁、镉、锰[J].中国无机分析化学(Chinese Journal of Inorganic Analytical Chemistry),2013,3(4):45-48.
[7]
Hammer M R. A magnetically excited microwave plasma source for atomic emission spectroscopy with performance approaching that of the inductively coupled plasma [J]. Spectrochimica Acta, 2008, 63B: 456-464.
[8]
陈颖,李承跃,季天仁. 新型大气压微波等离子体炬的仿真研究[J]. 强激光与粒子束(High Power Laser and Particle Beams). 2011,23(10):2715-2718.
[9]
赵丽巍,付瑶,宋大千,等.微波等离子体炬耦合方式的研究 [J]. 高等学校化学学报(Chemical Journal of Chinese Universities). 2001,22(2):201-205.