电感耦合等离子体原子发射光谱法测定铜铬锆合金中铬和锆
马琪, 李正明, 翟博
兵器工业西北地区理化检测中心,陕西西安 710043
Determination of chromium and zirconium in copper-chromium-zirconium alloy by inductively coupled plasma atomic emission spectrometry
MA Qi,LI Zheng-ming,ZHAI Bo
Ordnance Industry Northwestern Physical & Chemistry Testing Center, Xian 710043, China
摘要 以硝酸和硫酸溶解样品,通过选择元素间没有干扰的光谱线284.984 nm和 343.823 nm分别作为Cr和Zr分析线和采用铜基体匹配技术,用电感耦合等离子体原子发射光谱法(ICP-AES)测定了铜铬锆合金中铬和锆。铬和锆的检出限分别为0.009、0.012 μg/mL。方法用于铜铬锆合金样品中铬和锆的测定,测定结果分别与过硫酸铵氧化容量法(JB/T9552.2-1999)和二甲酚橙光度法(JB/T9552.5-1999)的测定结果相符,10次平行测定的相对标准偏差分别为0.40%、0.28%。
关键词 :
电感耦合等离子体原子发射光谱法(ICP-AES) ,
铜铬锆合金 ,
铬 ,
锆
Abstract :The sample was dissolved with nitric acid and sulfuric acid. The spectral lines without interference among elements at 284.984 nm and 343.823 nm were selected as the analytical lines for chromium and zirconium respectively. Chromium and zirconium in copper-chromium-zirconium alloy were determined by inductively coupled plasma atomic emission spectrometry (ICP-AES) with copper matrix matching technique. The detection limit of chromium and zirconium was 0.009 and 0.012 μg/mL, respectively. The proposed method was applied to the determination of chromium and zirconium in copper-chromium-zirconium alloy samples. The results were consistent with those obtained by ammonium persulfate oxidization volumetric method (JB/T9552.2-1999) and xylenol orange spectrophotometry (JB/T9552.5-1999), and the relative standard deviation for ten times parallel determination was 0.40 % and 0.28 %, respectively.
Key words :
inductively coupled plasma atomic emission spectrometry (ICP-AES)
copper-chromium-zirconium alloy
chromium
zirconium
收稿日期: 2013-07-02
作者简介 : 马 琪(1981-),工程师,从事化学分析工作;E-mail: huashanmq@163.com
[1] 王华炜.异步牵引电机用锆铬铜线中锆的测定[J].电力机车与城轨车辆(Electric Locomotives & Mass Transit Vehicles),2003,26(3):51-52.
[2] 顾继红,杨丽荣,于媛君.ICP-AES法同时分析有色金属铜及其合金中15种元素[J].冶金分析(Metallurgical Analysis),2004,24 (增):79-82.
[3] 李韶梅,石毓霞,赵伟.电感耦合等离子体原子发射光谱法测定镍硅铜中镍和铜[J].冶金分析(Metallurgical Analysis),2007,27(2):79-80.
[4] 游玉萍,熊晓燕,岳伟,等.电解分离-电感耦合等离子体原子发射光谱法测定砷铜合金中磷[J].冶金分析(Metallurgical Analysis),2008,28(9):51-53.
[5] 吴伟明,许方. ICP-AES法测定铜合金中的镍、铅、锡 [J].现代科学仪器(Modern Scientific Instruments),2008, (5):54-55.
[6] 李元之,刘沛.电感耦合等离子体原子发射光谱法测定铜合金中磷[J].理化检验-化学分册(Physical Testing and Chemical Analysis Part B:Chemical Analysis),2009,46(2):195-196,230.
[7] 刘洁,房书梅,张建忠,等.ICP-AES法测定铜合金中的Co、Cr、Zr元素含量[J].河北冶金(Hebei Metallurgy),2011,(11):62-63.
[8] 谢芬.ICP-AES法测定磷铜合金中的磷[J].武钢技术(Wuhan Iron and Steel Corporation Technology),2001,49(4):27-29.
[9] 张红侠,吴琴,徐婷婷,等.ICP-AES同时测定铜合金中Sn、Pb、Fe、Zn元素[J].山东科学(Shandong Science),2012,25(5):18-21.
[10] 马琪,蒙娟娟.电感耦合等离子体原子发射光谱法测定钼铜合金中镍、钙、镁、铝含量[J].理化检验-化学分册(Physical Testing and Chemical Analysis Part B:Chemical Analysis),2012,48(12):1496-1497.
[1]
赵宏伟, 吴鉴, 黄海艺, 谢克强. 电感耦合等离子体原子发射光谱法测定电炉熔炼黄磷电尘灰中镓 [J]. 冶金分析, 2019, 39(2): 56-60.
[2]
崔宏利. 熔融制样-X射线荧光光谱法测定硅质、铝质、镁质和铬质耐火材料中15种组分 [J]. 冶金分析, 2019, 39(2): 71-76.
[3]
冯晓军, 薛菁, 杨晓燕, 史鑫. 电感耦合等离子体原子发射光谱法测定云南昆阳磷矿黑色页岩中钒钼镍 [J]. 冶金分析, 2018, 38(9): 53-58.
[4]
李秋莹, 甘建壮, 何姣, 方海燕, 孙祺, 王应进. 多元光谱拟合校正-电感耦合等离子体原子发射光谱法测定铱化合物中19种杂质元素 [J]. 冶金分析, 2018, 38(9): 63-68.
[5]
张秀艳, 常诚, 赵静. 电感耦合等离子体原子发射光谱法测定铈铁合金中铝硅镍 [J]. 冶金分析, 2018, 38(9): 69-74.
[6]
王小欢. 便携式X射线荧光光谱仪分析镀锌板铬盐钝化膜涂镀量 [J]. 冶金分析, 2018, 38(8): 11-15.
[7]
胡维铸, 唐语, 燕际军. 直流辉光放电原子发射光谱法测定中低合金钢中锆 [J]. 冶金分析, 2018, 38(8): 16-20.
[8]
张志刚, 章发, 李方军, 张霄霄, 张勇. 电感耦合等离子体原子发射光谱法测定硅铁钡孕育剂中9种元素 [J]. 冶金分析, 2018, 38(8): 53-57.
[9]
陈丽梅, 陈兰, 罗正波, 蒋朝金, 黄前军. 火试金-电感耦合等离子体原子发射光谱法测定提钯螯合剂中残余钯 [J]. 冶金分析, 2018, 38(8): 58-62.
[10]
成勇. 电感耦合等离子体原子发射光谱法测定V-4Cr-4Ti合金中铬和钛 [J]. 冶金分析, 2018, 38(8): 63-69.
[11]
蔺菲, 王蓬, 李朝, 李冬玲, 赵雷. 惰气熔融-红外吸收/热导法测定铜铬合金中氧和氮 [J]. 冶金分析, 2018, 38(7): 38-43.
[12]
于媛君, 亢德华, 杨丽荣, 高品, 王莹, 邓军华. 电感耦合等离子体原子发射光谱法测定钢中钨钼铌方法标准的探讨 [J]. 冶金分析, 2018, 38(7): 73-79.
[13]
陈吉祥,杨树泉,程林. 电感耦合等离子体原子发射光谱法测定再生锌原料中锗 [J]. 冶金分析, 2018, 38(6): 43-49.
[14]
肖凡,张宁,姜云军,郭秀平,王趁荣,刘璐. 密闭酸溶-电感耦合等离子体原子发射光谱法测定地球化学调查样品中硼 [J]. 冶金分析, 2018, 38(6): 50-54.
[15]
宋祖峰, 陆向东, 荚江霞, 陆尹, 王昔文, 贾云海. 电感耦合等离子体原子发射光谱法测定低碳低钛硅铁中痕量铌钒锆 [J]. 冶金分析, 2018, 38(5): 35-40.