Spectrophotometric determination of microcadmium in water after ionic liquid extraction
LI Shan 1, CHEN Zhen 1, WANG Qian 2, WANG Yinghua* 1
1 College of Chemistry, Jilin University, Changchun 130012, China; 2 College of Materials Science and Engineering, Jilin University, Changchun 130012, China
Abstract: A new extraction spectrophotometry for the determination of microcadmium was proposed. In Na2B4O7NaOH buffer medium at pH 10.8, the CdCadion complex formed by 1(4 nitrobenzophenone)3(4phenylazophenyl) triazene (Cadion) and Cd2+ could be extracted by little ionic liquid (1butyl3methylimidazole hexafluorophosphate, [Bmim][PF6]). After centrifugal separation, the content of cadmium in ionic liquid phase was directly determined by spectrophotometry at maximum absorption wavelength (517 nm). The linear range of this method was 1.0-62.5 ng/mL, the detection limit was 0.22 ng/mL, apparent molar absorptivity was 6.9×105 L·mol-1·cm-1, the enhancement factor was 29.5, and the relative standard deviation (RSD, 50 ng/mL, n=6) was 2.1 %. The proposed method was directly applied to the determination of microcadmium in water sample with recoveries of 96.7 %-108.5 %.
李姗,陈珍,王千,王英华*. 离子液体萃取光度法测定水中微量镉[J]. 冶金分析, 2012, 32(02): 28-31.
LI Shan, CHEN Zhen, WANG Qian, WANG Yinghua*. Spectrophotometric determination of microcadmium in water after ionic liquid extraction. , 2012, 32(02): 28-31.