Abstract:Twenty-three trace impurities in high purity antimony such as Mg, Al, Si, Ca and Fe were determined by dynamic reaction cell inductively coupled plasma mass spectrometry. The determination isotope was optimized and selected. The spectral interference of Si, Ca and Fe was eliminated with H2 as reaction gas of dynamic reaction cell. The optimal H2 flow rate was established as 4 mL/min. The instrumental signal shift and matrix effect were compensated with Rh as internal standard element. The lower limits of quantitation were in the range of 0.037-0.77 ng/mL, the recoveries of standard addition were 92 %-112 %, and the relative standard deviations (RSD) were smaller than 10 %. This proposed method could meet the requirements for the determination of twenty-three trace impurities in 4N5 high purity antimony.
李继东,王长华,郑永章. 动态反应池-电感耦合等离子体质谱法测定高纯锑中23种痕量杂质元素[J]. 冶金分析, 2011, 31(12): 21-25.
LI Ji-dong, WANG Chang-hua, ZHENG Yong-zhang. Determination of twenty-three trace impurity elements in high purity antimony by dynamic reaction cell inductively coupled plasma mass spectrometry. , 2011, 31(12): 21-25.