Determination of LOD and LOQ with relative standard deviation and relative range or ratio of maximum and minimum
JIA Yunhai1, SUN Xiaofei2, ZHANG Fan3
1. Central Iron & Steel Research Institute,Beijing 100081,China; 2. NCS Testing Technology Co.,Ltd.,Beijing 100081,China; 3. China National Institute of Standardization,Beijing 100191,China
Abstract:The corresponding relative standard deviations (RSD) of limit of detection (LOD) and limit of quantification (LOQ) constitute two critical values,i.e.,RSDLOD(MAX) and RSDLOQ(MAX).In general,ten times and three times of standard deviation of the blank sample in 10 measurements are used as LOQ and LOD,respectively.At 95% confidence interval,the obtained RSD at concentration of LOQ and LOD are often larger than the theoretical value 10% or 33% of normal distribution.The McKay model for standard deviation of confidence interval was modified by Vangel.The critical value formula of RSD of small sample (K) was estimated from the RSD of normal distribution (k).It was found that when the LOQ and LOD was 10 times and 3.3 times of standard deviation of sample blank,the corresponding critical value of RSD for LOQ and LOD in ten measurements was 14% and 43%,respectively.The quantitative analysis was not available when RSD was larger than 14%,and it was undetectable when RSD was larger than 43%.When relative range (RR) was used to evaluate the precision,the corresponding critical value of RR for LOQ and LOD in two independent measurements was 28% and 84%,respectively.The quantitative analysis was not available when RR was larger than 28%,and it was undetectable when RSD was larger than 84%.The ratio of maximum and minimum (xmax/xmin) in 2~4 measurements could be used to provide the corresponding critical value for LOQ and LOD.The quantitative analysis was available when the ratio (x2/x1) was less than 1.3 in two independent measurements.The qualitative analysis was available when the ratio (x2/x1) was in range of 1.3<(x2/x1)≤2.5,while the qualitative analysis was not available when the ratio (x2/x1) was larger than 2.5.If LOQ and LOD was defined as 14.1 times and 4.65 times of standard deviation of sample blank (SB),the corresponding critical value of RSD for LOQ and LOD in ten measurements of blank sample was 10% and 30%,respectively.The quantitative analysis was not available when RSD was larger than 10%,and it was undetectable when RSD was larger than 30%.The corresponding critical value of RR for LOQ and LOD in two independent measurements was 20% and 61%,respectively.The quantitative analysis was not available when RSD was larger than 20%,and it was undetectable when RSD was larger than 61%.The quantitative analysis was available when the ratio (x2/x1) in two independent measurements was less than 1.2.The qualitative analysis was available when the ratio (x2/x1) was in range of 1.2<(x2/x1)≤1.9,while the qualitative analysis was not available when the ratio (x2/x1) was larger than 1.9.Different definition of LOD and LOQ led to different critical values of RSD and RR and (xmax/xmin).For any detection method and any testing concentration,when the RSD and number of measurements were provided,the use of RSD,RR or (xmax/xmin) ratio could be used to rapidly judge whether effective qualitative analysis or quantitative analysis could be conducted.
贾云海, 孙晓飞, 张帆. 用相对标准偏差和相对极差及测量极值比例确定分析检出限和定量限[J]. 冶金分析, 2021, 41(1): 1-12.
JIA Yunhai, SUN Xiaofei, ZHANG Fan. Determination of LOD and LOQ with relative standard deviation and relative range or ratio of maximum and minimum. , 2021, 41(1): 1-12.
Lloyd A Currie.Nomenclature in evaluation of analytical methods including detection and quantification capability [J].Pure &Apply.Chem.,1995,67(10):1699-1723.
[2]
V Barwick,P P M Bravo,S L R Ellison.The fitness for purpose of analytical methods:a laboratory guide to method validation and related topics[M].2da Edicion.B Magnusson,U Ornemark Eds,2014.
[3]
Gerhard Schwarz,Stefan Baumler,Annette Block,et al.Determination of detection and quantification limits for SNP allele frequency estimation for DNA pools using real time PCR [J].Nucleic Acids Research,2004,32(3):24.
ISO 13933-2014 Steel and iron Determination of calcium and magnesium-Inductively coupled plasma atomic emission spectrometric method[S].International Organization for Standardization,2014.
ASTM D 6061-07 Standard practice for 99%/95% Interlaboratory Detection Estimate (IDE) for analytical methods with negligible calibration error[S].ASTM International,2007.
[8]
李玉武,任立军,王婧瑞,等.方法检出限三个评估方法与比较[J].中国无机分析化学,2015,5(3):24-33.LI Yuwu,REN Lijun,WANG Jingrui,et al.Interpretation and comparison of three evaluation methods of method detection limit[J].Chinese Journal of Inorganic Analytical Chemistry,2015,5(3):24-33.
[9]
CHOI,Dong Hwan,KATAKURA,et al.Validation of a method for predicting the precision,limit of detection range of quantitation in competitive ELISA[J].Analytical Science,2007,23:215-218.
[10]
McKay A T.Distribution of the coefficient of variation and the extended ‘t’ distribution[J].Journal of the Royal Statistical Society,1932,95:695-698.
[11]
Vangel M.Confidence interval for normal coefficient of variation[J].American Statistician,1996(50):21-26.
[12]
Krishnamoorthy K,Meesook Lee.Improved tests for the equality of normal coefficients of variation [J].Computational Statistics,2013,28:3.
ASTM E1601-12 Standard practice for conducting an inter-laboratory study to evaluate the performance of an analytical method[S].ASTM International,2012.
[16]
ISO/TR 21074:2016(E) Application of ISO 5725 for the determination of repeatability and reproducibility of precision tests performed in standardization work for chemical analysis of steel[S].International Organization for Standardization,2016.
[17]
ISO/TC102/TCR12-Technical Committee Report No.12.N 587E [S].International Organization for Standardization,2007.
[18]
Thompson M.Recent trends in inter-laboratory precision at ppb and sub-ppb concentrations in relation to fitness for purpose criteria in proficiency testing [J].Analyst,2000,125(3):385-386.
[19]
Thompson M.An emergent optimal precision in chemical measurement at low concentrations [J].Analytical Methods,2013(5):4518-4519.
岳振峰,张志旭.日本食品中农药残留等检测方法评价指南[J].食品安全质量检测学报,2014,5(2):323-327.YUE Zhenfeng,ZHANG Zhixu.Japan guidance for the evaluation on the validity of the method for testing residues of pesticides and other substance in food[J].Journal of Food Safety and Quality,2014,5(2):323-327.
[22]
International Organization for Standardization.ISO TR 9769-2018 Steel and Iron-Review of available methods of analysis[S].2018.
ASTM F1710-2008(2016) Standard test method for trace metallic impurities in electronic grade titanium by high mass-resolution glow discharge mass spectrometer[S].ASTM International,2016.
[25]
ASTM F1593-2008(2016) Standard test method for trace metallic impurities in electronic grade aluminum by high mass-resolution glow-discharge mass spectrometer[S].ASTM International,2016.
[26]
ASTM F2405-2004(2011) Standard test method for trace metallic impurities in high purity copper by high-mass-resolution glow discharge mass spectrometer[S].ASTM International,2011.