Abstract:Standard gas calibration equipment for the simultaneous determination of O, N, H and Ar was built by combining the gas circuit of pulse heating and time-of-flight mass spectrometer gas analyzer with the six-port injection valve which was normally used in chromatogram. Two processes of purging of quantitative tube and injecting of standard gas can be converted smoothly. Using the equipment, calibration curves of low content O, N, H and Ar were established, and thestability was tested. Samples were analyzed by these calibration curves, and the results matched with the certified values. From the results, it was showed that the problem of Ar reference materials in ultra low content range can be solved by this equipment, and Ar can be analyzed quantitatively. The established calibrating curve by standard gas calibration setup can be used for metal materials analyzing, and good linearization of the calibrating curve can be got.
杨植岗,王学华,胡少成,,徐井然,吴振宁,沈学静,王 蓬,. 脉冲熔融-飞行时间质谱法分析痕量气体成分的校准方法研究[J]. 冶金分析, 2012, 32(10): 1-6.
YANG Zhi-gang,,WANG Xue-hua,,HU Shao-cheng,,XU Jing-ran,WU Zhen-ning,SHEN Xue-jing,,WANG Peng,. Research on the calibration method of pulse heating and time-of-flight mass spectrometric determination of trace gases. , 2012, 32(10): 1-6.
[1] 王蓬,沈学静,胡少成,等.脉冲熔融-飞行时间质谱仪-一种新型的冶金材料气体分析仪[C].2008年国际冶金及材料分析测试学术报告会论文集(International Conference and Exhibition on Analysis & Testing of Materials Proceedings),No.1:109-119.
[2] 高伟,胡少成,黄正旭,等.用于气体快速分析的飞行时间质谱仪的研制[C].2008年国际冶金及材料分析测试学术报告会论文集(International Conference and Exhibition on Analysis & Testing of Materials Proceedings), No.2:966-970.