Abstract: To overcome the difficults of the analysis of inert gas argon and the scarcity of argon reference materials, a kind of standard gas calibration device was designed and quantitative pipes of different volumes were prepared to conduct calibration by feeding standard gas through standard gas injection interface. Simultaneously, a novel determination method of argon in titanium alloy and nano-alloy powder was established by pulse inert gas fusion-mass spectrometry. The results indicated that the absolute quality of argon showed good linear relationship with ion count of Ar+(m/z = 40)spectral line. The linear range was 0.000 010%-0.10% and the detection limit was 0.000 006%. This proposed method was applied to the determination of argon in titanium alloy and nano-alloy powder. The results were consistent with those obtained by pulse fusion thermal conductivity method or theoretical calculated value, and the relative standard deviations were all less than 14%.
胡少成,杨植岗,郝贵奇,张淑芳,王 蓬,沈学静,马红权,王海舟 . 脉冲惰气熔融-质谱法测定钛合金及纳米合金粉中的氩[J]. 冶金分析, 2012, 32(6): 1-4.
HU Shao-cheng, YANG Zhi-gang, HAO Gui-qi, et al. Determination of argon in titanium alloy and nano-alloy powder by pulse insert gas fusion-mass spectrometry. , 2012, 32(6): 1-4.